*** MIDAS logfile (05SEPpl1.2) *** Su, 25 Nov 2007 15:05:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 25 Nov 2007 15:05:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 071125 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.11 stddev = 2.87 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_071125 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1918.7 1918.1 678.4 0.00 -0.01 2.83 2 8.63 3771.5 3771.5 1331.8 0.03 0.04 2.83 3 17.25 7466.7 7465.5 2623.6 0.01 0.00 2.85 4 25.87 11157.9 11158.8 4014.1 -0.03 -0.01 2.78 5 38.80 16700.3 16701.6 5893.6 -0.02 -0.01 2.83 6 51.73 22241.7 22239.4 7875.0 -0.02 -0.03 2.82 7 64.66 27784.9 27784.2 9958.0 -0.01 -0.02 2.79 8 86.21 37022.6 37020.7 13072.4 -0.01 -0.01 2.83 9 107.76 46274.1 46273.6 16300.8 0.02 0.02 2.84 10 129.31 55520.3 55519.2 19568.0 0.03 0.03 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5443 to 157536 e- Average count rate measured: 429 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.01 e-/ADU RON = 8.16 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.165 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay