*** MIDAS logfile (05SEPpl1.2) *** Su, 14 Oct 2007 13:58:17 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 Oct 2007 13:58:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 071014 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 560.74 stddev = 2.45 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_071014 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1927.1 1930.1 685.0 -0.03 -0.00 2.81 2 8.63 3793.0 3792.7 1362.9 0.08 0.02 2.78 3 17.25 7508.0 7506.6 2642.3 0.03 -0.00 2.84 4 25.87 11221.0 11217.4 3966.3 -0.00 -0.04 2.83 5 38.80 16790.4 16791.9 5957.7 -0.03 -0.01 2.82 6 51.73 22360.9 22360.6 7864.0 -0.04 -0.02 2.84 7 64.66 27932.9 27933.4 9815.1 -0.03 -0.02 2.85 8 86.21 37227.9 37226.3 13080.5 -0.01 0.00 2.85 9 107.76 46526.7 46528.1 16695.2 0.01 0.03 2.79 10 129.31 55823.4 55817.7 19413.9 0.02 0.03 2.88 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5488 to 158848 e- Average count rate measured: 431 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.02 e-/ADU RON = 6.96 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.165 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay