*** MIDAS logfile (05SEPpl1.2) *** Su, 16 Sep 2007 20:19:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 16 Sep 2007 20:19:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070916 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.31 stddev = 2.47 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070916 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1942.9 1943.3 691.2 -0.03 -0.00 2.81 2 8.63 3824.1 3821.4 1360.1 0.09 0.02 2.81 3 17.25 7566.2 7567.7 2668.3 -0.01 0.01 2.84 4 25.87 11313.1 11313.4 3988.0 0.00 0.01 2.84 5 38.80 16929.1 16926.6 6002.8 -0.02 -0.03 2.82 6 51.73 22543.7 22545.4 8058.7 -0.03 -0.02 2.80 7 64.66 28163.2 28163.9 9942.3 -0.02 -0.02 2.83 8 86.21 37530.4 37530.8 13129.9 -0.01 -0.01 2.86 9 107.76 46905.3 46905.0 16756.7 0.01 0.01 2.80 10 129.31 56280.4 56276.2 19796.0 0.03 0.02 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5503 to 159394 e- Average count rate measured: 435 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 7.01 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.162 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay