*** MIDAS logfile (05SEPpl1.2) *** Fr, 13 Jul 2007 17:14:44 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 13 Jul 2007 17:14:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070713 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 559.90 stddev = 3.84 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070713 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1973.5 1974.6 716.8 -0.02 -0.00 2.75 2 8.63 3878.6 3878.4 1388.1 0.03 0.01 2.79 3 17.25 7676.3 7676.0 2728.8 -0.00 -0.01 2.81 4 25.87 11476.1 11476.7 4041.8 0.01 0.01 2.84 5 38.80 17170.1 17169.8 6109.3 -0.02 -0.02 2.81 6 51.73 22865.3 22867.1 8120.9 -0.03 -0.02 2.82 7 64.66 28569.4 28565.3 10044.5 0.00 -0.01 2.84 8 86.21 38063.6 38062.4 13464.9 -0.01 -0.01 2.83 9 107.76 47572.6 47572.7 16824.8 0.02 0.03 2.83 10 129.31 57067.8 57064.7 20100.1 0.01 0.01 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5601 to 161925 e- Average count rate measured: 441 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.00 e-/ADU RON = 10.88 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.170 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay