*** MIDAS logfile (05SEPpl1.2) *** Sa, 07 Jul 2007 17:23:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 07 Jul 2007 17:24:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070707 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.14 stddev = 3.62 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1977.6 1979.6 712.8 0.00 0.03 2.77 2 8.63 3884.9 3885.2 1352.0 -0.01 -0.03 2.87 3 17.25 7693.9 7691.9 2713.8 0.03 -0.01 2.84 4 25.87 11496.1 11495.7 4097.5 -0.02 -0.03 2.81 5 38.80 17205.6 17205.1 6110.9 -0.02 -0.02 2.82 6 51.73 22912.4 22912.0 8091.1 -0.03 -0.02 2.83 7 64.66 28624.7 28623.6 10067.7 -0.01 -0.01 2.84 8 86.21 38143.1 38139.9 13378.0 -0.01 -0.00 2.85 9 107.76 47680.9 47674.6 16505.4 0.04 0.04 2.89 10 129.31 57197.5 57197.8 20277.2 0.03 0.05 2.82 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5639 to 163023 e- Average count rate measured: 442 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.02 e-/ADU RON = 10.32 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.170 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay