*** MIDAS logfile (05SEPpl1.2) *** Su, 13 May 2007 19:00:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 13 May 2007 19:00:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070513 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.37 stddev = 2.49 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070513 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2014.4 2014.7 717.6 0.01 0.02 2.81 2 8.63 3957.9 3958.2 1417.2 -0.02 -0.01 2.79 3 17.25 7838.3 7838.2 2749.1 -0.00 -0.00 2.85 4 25.87 11715.3 11716.5 4163.0 -0.02 -0.02 2.81 5 38.80 17534.2 17534.6 6299.2 -0.02 -0.02 2.78 6 51.73 23352.4 23351.8 8259.2 -0.02 -0.03 2.83 7 64.66 29172.2 29175.8 10328.6 -0.02 -0.01 2.82 8 86.21 38879.4 38878.7 13702.9 0.01 0.00 2.84 9 107.76 48595.5 48594.2 17368.3 0.04 0.03 2.80 10 129.31 58303.9 58298.6 20502.3 0.05 0.04 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5702 to 165009 e- Average count rate measured: 450 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 7.04 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.165 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay