*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 15:55:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 15:55:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.55 stddev = 2.43 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2002.6 2000.0 702.3 0.00 -0.02 2.85 2 8.63 3934.3 3934.2 1404.4 0.01 0.06 2.80 3 17.25 7788.4 7784.8 2748.8 0.01 -0.02 2.83 4 25.87 11637.9 11638.6 4102.2 -0.03 -0.01 2.84 5 38.80 17419.0 17416.4 6182.1 -0.02 -0.03 2.82 6 51.73 23196.5 23196.3 8152.6 -0.03 -0.03 2.85 7 64.66 28980.0 28980.1 10231.7 -0.01 -0.01 2.83 8 86.21 38617.0 38619.1 13622.4 -0.01 0.00 2.83 9 107.76 48273.4 48264.7 16992.5 0.04 0.02 2.84 10 129.31 57912.3 57911.3 20379.0 0.04 0.04 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5688 to 164586 e- Average count rate measured: 447 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.00 e-/ADU RON = 6.89 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.166 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay