*** MIDAS logfile (05SEPpl1.2) *** Fr, 16 Feb 2007 16:27:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 16 Feb 2007 16:28:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070216 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.92 stddev = 2.42 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2049.0 2048.9 725.0 -0.00 0.01 2.83 2 8.63 4027.1 4026.2 1431.3 0.03 0.02 2.81 3 17.25 7968.7 7968.5 2803.0 -0.02 -0.02 2.84 4 25.87 11912.3 11912.4 4190.3 -0.02 -0.02 2.84 5 38.80 17827.2 17827.5 6342.8 -0.02 -0.02 2.81 6 51.73 23743.0 23743.1 8311.9 -0.02 -0.02 2.86 7 64.66 29657.8 29660.7 10471.4 -0.02 -0.01 2.83 8 86.21 39527.0 39526.1 13865.9 0.00 -0.00 2.85 9 107.76 49400.7 49398.9 17411.8 0.03 0.02 2.84 10 129.31 59277.2 59276.7 20824.8 0.05 0.05 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5831 to 168691 e- Average count rate measured: 458 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 6.88 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.167 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay