*** MIDAS logfile (05SEPpl1.2) *** Sa, 20 Jan 2007 16:05:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 20 Jan 2007 16:05:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070120 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.66 stddev = 2.45 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_070120 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2042.6 2040.8 729.6 0.03 0.01 2.80 2 8.63 4009.3 4010.1 1413.2 -0.05 0.00 2.84 3 17.25 7940.4 7937.7 2800.3 -0.01 -0.03 2.84 4 25.87 11867.9 11867.1 4097.8 -0.02 -0.02 2.90 5 38.80 17763.9 17761.9 6248.6 -0.01 -0.01 2.84 6 51.73 23656.7 23654.1 8297.5 -0.01 -0.02 2.85 7 64.66 29551.0 29550.4 10231.4 -0.01 -0.01 2.89 8 86.21 39376.5 39378.1 13945.3 -0.01 0.00 2.82 9 107.76 49223.6 49221.1 17690.2 0.04 0.04 2.78 10 129.31 59052.9 59049.0 20458.7 0.04 0.04 2.89 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5806 to 167930 e- Average count rate measured: 456 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.03 e-/ADU RON = 6.98 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.167 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay