*** MIDAS logfile (05SEPpl1.2) *** Fr, 22 Dec 2006 19:50:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 22 Dec 2006 19:50:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061222 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 565.42 stddev = 2.43 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061222 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2079.5 2079.2 730.3 0.04 0.04 2.85 2 8.63 4083.2 4082.9 1436.4 -0.04 -0.04 2.84 3 17.25 8085.2 8084.1 2852.8 -0.02 -0.03 2.83 4 25.87 12085.4 12087.5 4228.4 -0.03 -0.02 2.86 5 38.80 18088.1 18090.6 6382.6 -0.03 -0.02 2.83 6 51.73 24091.9 24093.2 8359.8 -0.02 -0.02 2.88 7 64.66 30097.1 30096.0 10593.2 -0.01 -0.02 2.84 8 86.21 40110.5 40111.3 14299.3 0.01 0.01 2.81 9 107.76 50135.6 50130.6 17448.0 0.05 0.03 2.87 10 129.31 60152.0 60152.4 20952.0 0.06 0.05 2.87 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5954 to 172228 e- Average count rate measured: 464 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.02 e-/ADU RON = 6.96 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.166 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay