*** MIDAS logfile (05SEPpl1.2) *** Th, 14 Dec 2006 18:21:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 14 Dec 2006 18:21:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061214 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.92 stddev = 2.49 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061214 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2068.7 2068.9 736.8 -0.02 -0.03 2.81 2 8.63 4069.8 4069.3 1454.3 0.07 0.05 2.80 3 17.25 8055.1 8057.0 2817.1 0.02 0.04 2.86 4 25.87 12038.2 12038.7 4267.5 -0.02 -0.01 2.82 5 38.80 18011.7 18014.4 6406.8 -0.05 -0.03 2.81 6 51.73 23994.9 23992.4 8372.1 -0.02 -0.03 2.87 7 64.66 29970.6 29974.5 10622.9 -0.03 -0.02 2.82 8 86.21 39944.9 39945.1 14071.0 -0.00 -0.00 2.84 9 107.76 49926.2 49923.3 17527.2 0.03 0.02 2.85 10 129.31 59896.2 59892.9 20655.2 0.03 0.02 2.90 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5957 to 172463 e- Average count rate measured: 462 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.88 +/- 0.02 e-/ADU RON = 7.16 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.164 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay