*** MIDAS logfile (05SEPpl1.2) *** Fr, 01 Dec 2006 14:28:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 01 Dec 2006 14:28:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061201 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.11 stddev = 2.53 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2088.1 2091.2 736.1 -0.02 0.01 2.84 2 8.63 4109.4 4108.1 1450.7 0.06 -0.02 2.83 3 17.25 8135.5 8135.9 2842.8 0.01 -0.00 2.86 4 25.87 12159.4 12161.6 4251.7 -0.03 -0.01 2.86 5 38.80 18200.8 18199.7 6395.7 -0.02 -0.02 2.85 6 51.73 24239.5 24238.5 8518.3 -0.03 -0.02 2.85 7 64.66 30281.2 30281.7 10458.0 -0.02 -0.01 2.90 8 86.21 40354.0 40353.5 14278.6 -0.01 0.01 2.83 9 107.76 50438.8 50433.0 17665.4 0.03 0.03 2.86 10 129.31 60513.7 60507.2 21193.5 0.03 0.04 2.86 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5961 to 172609 e- Average count rate measured: 467 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 7.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.163 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay