*** MIDAS logfile (05SEPpl1.2) *** Sa, 25 Nov 2006 16:15:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 25 Nov 2006 16:16:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061125 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.16 stddev = 2.61 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061125 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2084.5 2085.7 743.2 0.01 0.01 2.80 2 8.63 4099.3 4099.2 1461.5 0.02 -0.00 2.80 3 17.25 8116.3 8117.6 2839.1 -0.01 0.00 2.86 4 25.87 12132.9 12132.5 4226.5 -0.02 -0.03 2.87 5 38.80 18160.1 18159.8 6426.2 -0.02 -0.02 2.83 6 51.73 24185.0 24184.4 8465.1 -0.03 -0.02 2.86 7 64.66 30215.2 30215.6 10694.2 -0.01 -0.01 2.83 8 86.21 40263.9 40265.0 14093.7 -0.01 0.00 2.86 9 107.76 50330.3 50323.5 17514.6 0.03 0.03 2.87 10 129.31 60380.7 60374.8 21154.3 0.04 0.03 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5968 to 172804 e- Average count rate measured: 466 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.01 e-/ADU RON = 7.47 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.162 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay