*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Nov 2006 12:55:33 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Nov 2006 12:55:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061117 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.61 stddev = 2.52 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061117 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2110.1 2109.5 746.7 0.03 0.00 2.83 2 8.63 4143.3 4144.6 1477.2 -0.05 -0.02 2.80 3 17.25 8207.6 8208.8 2913.3 -0.00 0.01 2.82 4 25.87 12270.2 12270.6 4363.3 0.00 0.00 2.81 5 38.80 18357.4 18358.7 6504.1 -0.03 -0.03 2.82 6 51.73 24455.0 24453.4 8574.5 -0.01 -0.02 2.85 7 64.66 30543.8 30548.1 10916.7 -0.02 -0.01 2.80 8 86.21 40708.7 40702.7 14095.0 0.00 -0.01 2.89 9 107.76 50881.7 50878.9 17754.5 0.04 0.03 2.87 10 129.31 61046.4 61039.0 21505.7 0.04 0.03 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6030 to 174461 e- Average count rate measured: 471 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.01 e-/ADU RON = 7.19 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.166 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay