*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Nov 2006 19:58:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Nov 2006 19:58:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061111 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.35 stddev = 2.55 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2079.2 2080.9 740.5 0.01 0.02 2.81 2 8.63 4087.3 4088.5 1460.3 -0.01 -0.01 2.80 3 17.25 8097.5 8094.3 2860.4 0.02 -0.03 2.83 4 25.87 12103.0 12101.7 4232.1 -0.01 -0.02 2.86 5 38.80 18110.8 18110.9 6261.1 -0.03 -0.02 2.89 6 51.73 24120.3 24121.8 8495.0 -0.03 -0.02 2.84 7 64.66 30133.4 30139.0 10543.9 -0.02 0.00 2.86 8 86.21 40161.8 40157.6 13924.9 0.00 0.00 2.88 9 107.76 50198.4 50192.6 17698.4 0.03 0.03 2.84 10 129.31 60222.9 60219.5 21131.6 0.04 0.04 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5932 to 171738 e- Average count rate measured: 465 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 7.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.163 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay