*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 13:22:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 13:23:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.95 stddev = 2.57 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2107.2 2109.0 758.3 0.00 0.01 2.78 2 8.63 4140.8 4140.4 1476.8 0.02 -0.02 2.80 3 17.25 8193.8 8198.5 2917.5 -0.03 0.02 2.81 4 25.87 12251.5 12247.5 4241.5 -0.01 -0.04 2.89 5 38.80 18330.6 18333.2 6461.2 -0.03 -0.02 2.84 6 51.73 24414.9 24416.2 8652.2 -0.02 -0.02 2.82 7 64.66 30501.2 30502.0 10971.1 -0.01 -0.01 2.78 8 86.21 40644.6 40642.6 14501.8 -0.00 -0.00 2.80 9 107.76 50809.6 50803.1 17945.7 0.05 0.04 2.83 10 129.31 60944.8 60946.4 21229.0 0.03 0.04 2.87 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6001 to 173502 e- Average count rate measured: 471 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.02 e-/ADU RON = 7.31 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.170 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay