*** MIDAS logfile (05SEPpl1.2) *** Fr, 15 Sep 2006 14:45:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 15 Sep 2006 14:46:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060915 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.45 stddev = 2.49 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060915 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2123.8 2126.7 756.1 -0.01 0.02 2.81 2 8.63 4176.6 4178.1 1468.0 0.01 -0.00 2.85 3 17.25 8272.7 8270.1 2952.5 0.02 -0.03 2.80 4 25.87 12364.5 12364.4 4352.6 -0.01 -0.02 2.84 5 38.80 18502.9 18503.1 6542.1 -0.03 -0.02 2.83 6 51.73 24643.4 24641.6 8662.6 -0.03 -0.03 2.84 7 64.66 30785.6 30783.9 10861.8 -0.02 -0.02 2.83 8 86.21 41032.0 41029.4 14381.9 0.00 0.01 2.85 9 107.76 51280.9 51279.5 18119.0 0.03 0.04 2.83 10 129.31 61525.0 61522.5 21496.3 0.03 0.04 2.86 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6065 to 175561 e- Average count rate measured: 475 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 7.11 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.164 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay