*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 20:28:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 20:29:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.78 stddev = 2.36 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2148.0 2148.6 772.2 -0.01 -0.00 2.78 2 8.63 4225.8 4226.5 1474.5 0.01 0.02 2.87 3 17.25 8372.1 8369.4 2947.8 0.03 -0.01 2.84 4 25.87 12511.9 12512.9 4418.4 -0.02 -0.01 2.83 5 38.80 18726.6 18727.4 6568.4 -0.02 -0.02 2.85 6 51.73 24941.2 24943.1 8782.3 -0.03 -0.02 2.84 7 64.66 31157.2 31160.6 10983.5 -0.02 -0.01 2.84 8 86.21 41524.4 41525.7 14620.5 -0.00 -0.00 2.84 9 107.76 51898.8 51903.3 18286.4 0.02 0.03 2.84 10 129.31 62269.0 62270.3 21846.1 0.03 0.04 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6113 to 177191 e- Average count rate measured: 481 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.00 e-/ADU RON = 6.70 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.158 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay