*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Aug 2006 16:18:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Aug 2006 16:19:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060810 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.87 stddev = 2.27 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060810 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2155.0 2157.0 769.1 0.00 0.03 2.80 2 8.63 4235.2 4233.8 1507.0 0.00 -0.05 2.81 3 17.25 8385.4 8384.1 2977.1 -0.01 -0.02 2.82 4 25.87 12535.4 12533.4 4421.2 -0.01 -0.02 2.84 5 38.80 18760.2 18757.7 6642.4 -0.02 -0.02 2.82 6 51.73 24984.9 24981.1 8915.9 -0.02 -0.02 2.80 7 64.66 31211.1 31209.5 10918.2 -0.02 -0.00 2.86 8 86.21 41592.1 41592.5 14600.9 -0.00 0.02 2.85 9 107.76 51988.0 51983.1 18124.6 0.03 0.04 2.87 10 129.31 62368.6 62368.8 21744.5 0.03 0.05 2.87 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6192 to 179130 e- Average count rate measured: 481 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.87 +/- 0.01 e-/ADU RON = 6.51 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.167 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay