*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 20:15:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 20:16:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.31 stddev = 2.31 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2161.9 2160.7 755.7 -0.02 -0.03 2.86 2 8.63 4252.9 4253.1 1532.0 0.05 0.08 2.78 3 17.25 8415.4 8416.7 2987.8 -0.02 -0.00 2.82 4 25.87 12582.5 12580.2 4449.5 -0.02 -0.03 2.83 5 38.80 18830.6 18830.8 6654.2 -0.02 -0.02 2.83 6 51.73 25080.1 25079.6 8884.3 -0.02 -0.02 2.82 7 64.66 31330.8 31330.2 10918.3 -0.02 -0.02 2.87 8 86.21 41753.6 41751.0 14773.6 0.00 -0.01 2.83 9 107.76 52184.2 52183.2 18376.4 0.03 0.02 2.84 10 129.31 62610.1 62607.4 21673.0 0.04 0.03 2.89 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6204 to 179728 e- Average count rate measured: 483 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.87 +/- 0.01 e-/ADU RON = 6.62 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.162 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay