*** MIDAS logfile (05SEPpl1.2) *** Fr, 07 Jul 2006 14:38:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 07 Jul 2006 14:39:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060707 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.28 stddev = 2.34 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2176.0 2179.1 775.7 -0.03 0.00 2.81 2 8.63 4289.1 4288.7 1517.7 0.07 0.02 2.83 3 17.25 8494.6 8492.0 2980.3 -0.00 -0.04 2.85 4 25.87 12700.5 12698.6 4530.0 -0.03 -0.04 2.80 5 38.80 19012.8 19009.3 6679.6 -0.02 -0.03 2.85 6 51.73 25321.2 25321.7 8875.5 -0.04 -0.02 2.85 7 64.66 31639.5 31639.2 11142.7 -0.01 0.00 2.84 8 86.21 42164.5 42168.8 14681.1 -0.00 0.02 2.87 9 107.76 52705.7 52704.4 18415.8 0.03 0.05 2.86 10 129.31 63233.7 63227.5 22172.8 0.04 0.05 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6232 to 180957 e- Average count rate measured: 488 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.01 e-/ADU RON = 6.68 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.151 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay