*** MIDAS logfile (05SEPpl1.2) *** Su, 28 May 2006 17:30:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 28 May 2006 17:31:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060528 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.60 stddev = 3.35 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060528 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2079.5 2083.0 746.0 -0.03 0.02 2.79 2 8.63 4100.5 4098.9 1490.0 0.08 -0.01 2.75 3 17.25 8121.1 8122.4 2933.3 -0.01 -0.01 2.77 4 25.87 12144.6 12146.0 4298.3 -0.02 -0.02 2.83 5 38.80 18182.2 18180.8 6437.8 -0.01 -0.02 2.82 6 51.73 24215.8 24215.3 8563.8 -0.03 -0.02 2.83 7 64.66 30254.7 30254.6 10669.4 -0.02 -0.01 2.84 8 86.21 40319.0 40319.6 14160.8 -0.01 0.00 2.85 9 107.76 50399.7 50396.1 17945.0 0.03 0.04 2.81 10 129.31 60458.9 60460.2 21005.5 0.02 0.04 2.88 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5950 to 172858 e- Average count rate measured: 467 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.02 e-/ADU RON = 9.59 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.148 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay