*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 20:44:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 20:45:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.79 stddev = 2.25 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 2129.9 2130.5 760.6 -0.02 0.02 2.80 2 8.63 4194.0 4189.8 1476.6 0.05 -0.03 2.84 3 17.25 8305.9 8304.9 2967.0 0.01 0.01 2.80 4 25.87 12416.3 12416.4 4346.0 -0.02 -0.01 2.86 5 38.80 18584.6 18582.6 6644.6 -0.03 -0.02 2.80 6 51.73 24754.4 24748.9 8723.2 -0.02 -0.03 2.84 7 64.66 30924.8 30920.7 10788.0 -0.02 -0.02 2.87 8 86.21 41212.0 41209.9 14635.0 -0.01 0.00 2.82 9 107.76 51512.9 51506.6 18095.4 0.03 0.03 2.85 10 129.31 61798.5 61799.0 21433.6 0.03 0.04 2.88 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6109 to 177220 e- Average count rate measured: 477 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.87 +/- 0.01 e-/ADU RON = 6.44 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.154 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay