*** MIDAS logfile (05SEPpl1.2) *** Mo, 19 Dec 2005 13:52:16 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 19 Dec 2005 13:53:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 051219 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.04 stddev = 2.35 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_051219 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1259.6 1261.8 450.7 -0.05 -0.01 2.79 2 8.63 2482.8 2482.6 886.7 0.08 0.01 2.80 3 17.25 4918.1 4917.8 1753.5 0.03 0.01 2.80 4 25.87 7353.5 7354.4 2614.8 0.02 0.03 2.81 5 38.80 11005.1 11005.5 3833.3 -0.01 0.00 2.87 6 51.73 14658.8 14655.5 5208.4 -0.00 -0.02 2.81 7 64.66 18307.8 18307.5 6471.3 -0.03 -0.02 2.83 8 86.21 24395.1 24395.4 8548.7 -0.03 -0.01 2.85 9 107.76 30488.2 30484.3 10765.5 -0.01 -0.01 2.83 10 129.31 36579.4 36576.9 12899.2 -0.00 0.01 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3580 to 103867 e- Average count rate measured: 283 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.01 e-/ADU RON = 6.67 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.153 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay