*** MIDAS logfile (03SEPpl1.2) *** Fr, 16 Sep 2005 18:01:42 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 16 Sep 2005 18:02:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 050916 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 562.88 stddev = 2.40 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_050916 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1296.2 1297.9 522.5 -0.03 -0.02 2.48 2 8.63 2554.5 2555.1 914.5 0.06 0.03 2.79 3 17.25 5061.1 5061.3 1823.1 0.03 0.02 2.78 4 25.87 7566.2 7566.7 2657.9 0.00 -0.00 2.85 5 38.80 11325.1 11326.2 3999.2 -0.01 0.00 2.83 6 51.73 15083.6 15085.7 5320.6 -0.01 0.00 2.83 7 64.66 18842.2 18843.8 6689.4 -0.02 -0.01 2.82 8 86.21 25108.3 25106.0 8825.4 -0.02 -0.02 2.84 9 107.76 31375.4 31374.5 11152.0 -0.01 -0.01 2.81 10 129.31 37646.9 37643.8 13161.3 0.00 0.01 2.86 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3698 to 107339 e- Average count rate measured: 291 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 6.85 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.152 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay