*** MIDAS logfile (03SEPpl1.2) *** Fr, 12 Aug 2005 13:54:11 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 12 Aug 2005 13:55:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 050812 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.43 stddev = 2.73 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_050812 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1293.5 1294.1 467.0 -0.01 0.01 2.77 2 8.63 2547.8 2546.3 928.5 0.03 -0.03 2.74 3 17.25 5049.2 5049.1 1806.8 0.03 0.03 2.79 4 25.87 7548.6 7549.2 2680.7 -0.00 0.02 2.82 5 38.80 11299.5 11298.8 3986.0 -0.00 0.00 2.83 6 51.73 15048.2 15047.7 5229.2 -0.02 -0.01 2.88 7 64.66 18800.4 18798.7 6651.8 -0.01 -0.01 2.83 8 86.21 25051.0 25048.5 8972.1 -0.01 -0.01 2.79 9 107.76 31305.7 31302.1 11041.1 -0.00 -0.00 2.84 10 129.31 37559.0 37557.1 13243.8 0.00 0.01 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3665 to 106405 e- Average count rate measured: 290 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 7.74 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.150 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay