*** MIDAS logfile (03SEPpl1.2) *** Sa, 16 Apr 2005 17:36:31 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 16 Apr 2005 17:37:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 050416 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.15 stddev = 2.41 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_050416 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1331.6 1332.5 478.5 -0.01 0.02 2.78 2 8.63 2622.5 2621.3 934.2 0.04 -0.03 2.81 3 17.25 5195.1 5196.2 1830.9 -0.00 0.01 2.84 4 25.87 7769.0 7770.4 2750.8 -0.00 0.01 2.82 5 38.80 11629.1 11629.6 4146.7 -0.01 -0.00 2.80 6 51.73 15488.7 15491.0 5504.3 -0.01 0.00 2.81 7 64.66 19348.1 19350.7 6889.6 -0.02 -0.00 2.81 8 86.21 25782.8 25780.8 9122.5 -0.01 -0.02 2.83 9 107.76 32222.4 32220.4 11318.1 0.01 0.00 2.85 10 129.31 38657.6 38658.6 13497.3 0.01 0.01 2.86 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3805 to 110430 e- Average count rate measured: 299 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.01 e-/ADU RON = 6.89 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.151 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay