*** MIDAS logfile (03SEPpl1.2) *** Su, 30 Jan 2005 16:08:06 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 30 Jan 2005 16:09:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 050130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.97 stddev = 2.51 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_050130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1346.8 1349.0 475.8 -0.02 0.02 2.83 2 8.63 2654.0 2654.2 938.6 0.03 -0.02 2.83 3 17.25 5260.7 5259.7 1852.6 0.02 -0.01 2.84 4 25.87 7866.5 7866.0 2769.9 0.01 -0.00 2.84 5 38.80 11776.7 11773.9 4158.0 0.01 -0.01 2.83 6 51.73 15682.1 15685.5 5834.6 -0.02 0.01 2.69 7 64.66 19592.6 19589.9 6909.0 -0.01 -0.01 2.84 8 86.21 26106.5 26107.5 9327.2 -0.02 0.00 2.80 9 107.76 32626.2 32623.3 11266.8 -0.00 0.01 2.90 10 129.31 39145.3 39141.7 13728.2 0.01 0.02 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3855 to 111965 e- Average count rate measured: 302 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.03 e-/ADU RON = 7.17 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.148 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay