*** MIDAS logfile (03SEPpl1.2) *** Fr, 19 Nov 2004 16:45:07 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 19 Nov 2004 16:46:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 041119 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.44 stddev = 3.86 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_041119 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1390.9 1389.8 510.5 -0.02 -0.06 2.72 2 8.63 2739.0 2740.0 993.3 0.04 0.10 2.76 3 17.25 5427.0 5425.9 1925.5 0.04 0.03 2.82 4 25.87 8111.2 8112.9 2896.1 -0.01 0.02 2.80 5 38.80 12141.8 12140.4 4310.2 -0.01 -0.02 2.82 6 51.73 16170.9 16170.1 5738.4 -0.01 -0.02 2.82 7 64.66 20201.1 20200.7 7063.2 -0.01 -0.02 2.86 8 86.21 26917.9 26917.0 9434.1 -0.01 -0.02 2.85 9 107.76 33639.2 33636.7 12183.7 0.00 -0.01 2.76 10 129.31 40355.2 40356.3 14116.6 -0.00 -0.00 2.86 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3938 to 114308 e- Average count rate measured: 312 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.02 e-/ADU RON = 10.94 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.152 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay