*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 28 Jun 2004 22:49:02 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 28 Jun 2004 22:50:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 040628 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.42 stddev = 2.28 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_040628 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1439.9 1439.5 515.1 -0.05 -0.03 2.80 2 8.63 2837.2 2835.1 1009.3 0.08 0.03 2.81 3 17.25 5618.7 5619.6 1999.9 0.02 0.05 2.81 4 25.87 8399.7 8399.3 2953.1 -0.00 0.00 2.84 5 38.80 12570.9 12573.1 4402.6 -0.02 0.00 2.86 6 51.73 16742.6 16742.7 5928.0 -0.03 -0.02 2.82 7 64.66 20916.7 20915.9 7360.1 -0.02 -0.02 2.84 8 86.21 27872.8 27871.0 9729.3 -0.01 -0.02 2.86 9 107.76 34835.2 34829.9 12161.7 0.01 -0.01 2.86 10 129.31 41793.8 41790.5 14700.5 0.01 0.01 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4112 to 119373 e- Average count rate measured: 323 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.86 +/- 0.01 e-/ADU RON = 6.51 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.152 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay