*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 14 Jun 2004 20:57:12 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 14 Jun 2004 20:58:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 040614 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.87 stddev = 3.65 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_040614 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1455.1 1452.9 525.0 0.03 -0.01 2.77 2 8.63 2861.0 2861.2 1039.9 -0.04 0.02 2.75 3 17.25 5669.5 5671.5 1979.6 -0.02 0.04 2.86 4 25.87 8479.5 8477.9 2961.4 0.01 -0.00 2.86 5 38.80 12690.2 12691.1 4542.6 -0.01 -0.00 2.79 6 51.73 16900.5 16901.4 6023.8 -0.02 -0.02 2.81 7 64.66 21114.9 21114.6 7487.7 -0.00 -0.02 2.82 8 86.21 28134.8 28137.5 9974.2 -0.01 -0.01 2.82 9 107.76 35163.7 35160.1 12455.4 0.02 -0.01 2.82 10 129.31 42191.3 42189.6 14873.3 0.03 0.01 2.84 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4116 to 119436 e- Average count rate measured: 326 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 10.34 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.152 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay