*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 19 Apr 2004 18:15:54 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 19 Apr 2004 18:17:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 040419 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 564.20 stddev = 2.23 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_040419 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1472.0 1469.8 526.5 0.01 -0.04 2.80 2 8.63 2895.2 2897.5 1034.2 -0.02 0.09 2.80 3 17.25 5738.0 5738.6 2051.4 0.01 0.03 2.80 4 25.87 8580.8 8579.7 2989.5 0.02 0.01 2.87 5 38.80 12838.5 12839.9 4546.3 -0.02 -0.02 2.82 6 51.73 17100.6 17100.5 5959.6 -0.02 -0.03 2.87 7 64.66 21362.9 21362.6 7564.6 -0.01 -0.03 2.82 8 86.21 28467.1 28467.5 9997.9 -0.01 -0.02 2.85 9 107.76 35579.9 35576.5 12608.3 0.02 -0.00 2.82 10 129.31 42682.7 42685.6 15169.2 0.01 0.01 2.81 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4149 to 120409 e- Average count rate measured: 330 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.82 +/- 0.01 e-/ADU RON = 6.30 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.152 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay