*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Jul 2003 17:51:12 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Jul 2003 17:51:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 030706 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.10 stddev = 2.28 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_030706 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1583.4 1584.1 565.7 -0.00 -0.02 2.80 2 8.63 3114.6 3116.9 1103.8 0.00 0.04 2.82 3 17.25 6168.7 6169.8 2213.5 -0.01 -0.01 2.79 4 25.87 9224.7 9224.8 3251.0 0.00 -0.01 2.84 5 38.80 13804.9 13806.7 4924.3 -0.01 -0.01 2.80 6 51.73 18388.4 18387.2 6544.6 -0.00 -0.01 2.81 7 64.66 22968.3 22970.4 8099.7 -0.01 -0.01 2.84 8 86.21 30608.1 30604.5 10730.5 -0.00 -0.02 2.85 9 107.76 38254.6 38250.2 13516.8 0.02 0.01 2.83 10 129.31 45894.4 45897.2 16327.6 0.02 0.03 2.81 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4475 to 129683 e- Average count rate measured: 354 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 6.44 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.159 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay