*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 05 Mar 2003 17:08:41 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 05 Mar 2003 17:09:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 030305 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 565.12 stddev = 2.36 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_030305 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1617.0 1616.7 583.7 -0.03 -0.02 2.77 2 8.63 3185.7 3185.2 1129.2 0.04 0.04 2.82 3 17.25 6313.2 6312.3 2309.8 0.03 0.03 2.73 4 25.87 9438.0 9437.4 3315.7 0.00 0.00 2.85 5 38.80 14126.0 14124.6 5014.2 -0.01 -0.02 2.82 6 51.73 18814.7 18814.6 6646.7 -0.02 -0.01 2.83 7 64.66 23502.4 23501.6 8187.3 -0.02 -0.02 2.87 8 86.21 31321.2 31320.6 10984.8 -0.01 -0.01 2.85 9 107.76 39141.7 39139.2 13485.2 0.00 0.00 2.90 10 129.31 46969.7 46971.2 16722.4 0.02 0.03 2.81 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4612 to 133979 e- Average count rate measured: 363 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.02 e-/ADU RON = 6.74 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.148 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay