*** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 26 Oct 2002 17:51:53 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 26 Oct 2002 17:52:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 021026 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.42 stddev = 2.32 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_021026 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1663.2 1663.5 596.5 -0.05 -0.02 2.79 2 8.63 3277.7 3275.8 1158.5 0.09 0.04 2.83 3 17.25 6488.1 6488.3 2318.3 0.00 0.00 2.80 4 25.87 9700.1 9703.4 3392.3 -0.02 0.02 2.86 5 38.80 14520.4 14518.9 5114.2 -0.02 -0.02 2.84 6 51.73 19341.1 19338.1 6833.4 -0.01 -0.02 2.83 7 64.66 24159.9 24160.2 8565.7 -0.02 -0.01 2.82 8 86.21 32196.9 32194.3 11337.9 -0.01 -0.01 2.84 9 107.76 40237.9 40237.2 13990.5 0.01 0.01 2.88 10 129.31 48278.4 48276.8 17044.8 0.02 0.02 2.83 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4739 to 137541 e- Average count rate measured: 373 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 6.60 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.153 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay