*** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 24 Oct 2002 14:03:43 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 24 Oct 2002 14:04:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 021024 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.21 stddev = 2.35 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_021024 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1665.7 1665.6 579.1 -0.03 0.01 2.88 2 8.63 3279.5 3276.8 1149.1 0.05 -0.01 2.85 3 17.25 6494.7 6494.2 2301.9 0.01 0.02 2.82 4 25.87 9711.4 9708.3 3397.5 0.01 -0.01 2.86 5 38.80 14532.0 14530.0 5079.3 -0.02 -0.02 2.86 6 51.73 19354.2 19351.5 6722.2 -0.02 -0.03 2.88 7 64.66 24176.5 24175.6 8507.6 -0.03 -0.02 2.84 8 86.21 32216.1 32221.3 11550.5 -0.02 0.00 2.79 9 107.76 40269.8 40268.2 14310.2 0.02 0.02 2.81 10 129.31 48317.9 48312.4 17088.2 0.03 0.03 2.83 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4686 to 135914 e- Average count rate measured: 373 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.81 +/- 0.01 e-/ADU RON = 6.60 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.155 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay