*** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 28 Sep 2002 16:31:49 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 28 Sep 2002 16:32:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020928 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.80 stddev = 2.50 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_020928 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1690.7 1692.2 612.6 0.00 -0.01 2.76 2 8.63 3327.0 3329.8 1183.2 -0.02 0.03 2.81 3 17.25 6594.1 6593.1 2301.3 0.01 -0.02 2.87 4 25.87 9860.2 9859.3 3511.4 0.00 -0.01 2.81 5 38.80 14757.7 14756.6 5295.7 -0.01 -0.01 2.79 6 51.73 19652.9 19653.5 6856.0 -0.03 -0.02 2.87 7 64.66 24556.2 24552.7 8577.3 -0.01 -0.01 2.86 8 86.21 32720.7 32720.0 11483.5 -0.01 0.00 2.85 9 107.76 40899.7 40896.0 14437.1 0.03 0.03 2.83 10 129.31 49067.5 49061.3 16900.4 0.03 0.03 2.90 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4873 to 141356 e- Average count rate measured: 379 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.88 +/- 0.02 e-/ADU RON = 7.21 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.154 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay