*** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 16 Aug 2002 16:19:11 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 16 Aug 2002 16:19:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020816 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.63 stddev = 2.41 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_020816 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1710.7 1709.3 617.2 0.00 -0.03 2.77 2 8.63 3367.9 3368.4 1184.0 0.03 0.06 2.84 3 17.25 6671.3 6672.0 2361.9 -0.01 0.01 2.82 4 25.87 9975.6 9976.3 3530.3 -0.01 0.00 2.83 5 38.80 14931.2 14932.2 5325.2 -0.02 -0.02 2.80 6 51.73 19889.2 19886.1 7086.9 -0.01 -0.03 2.81 7 64.66 24843.7 24848.2 8657.0 -0.02 -0.01 2.87 8 86.21 33110.3 33108.7 11551.0 0.00 -0.01 2.87 9 107.76 41383.2 41381.7 14655.2 0.02 0.01 2.82 10 129.31 49650.9 49647.7 17625.2 0.03 0.02 2.82 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4839 to 140495 e- Average count rate measured: 383 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.01 e-/ADU RON = 6.83 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.150 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay