*** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 02 Aug 2002 13:56:43 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 02 Aug 2002 13:57:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020802 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 562.49 stddev = 2.41 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_020802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1725.8 1726.0 607.8 -0.01 -0.01 2.84 2 8.63 3394.5 3395.3 1193.3 0.02 0.05 2.84 3 17.25 6723.2 6719.0 2378.9 0.03 -0.03 2.83 4 25.87 10046.1 10048.1 3529.5 -0.03 0.00 2.85 5 38.80 15037.1 15035.7 5324.7 -0.02 -0.02 2.82 6 51.73 20029.5 20026.3 7012.7 -0.01 -0.02 2.86 7 64.66 25022.3 25019.4 8877.1 0.00 -0.01 2.82 8 86.21 33339.5 33337.8 11765.8 -0.01 -0.01 2.83 9 107.76 41670.3 41671.0 14724.3 0.02 0.03 2.83 10 129.31 49989.4 49989.6 17788.7 0.02 0.03 2.81 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4863 to 140842 e- Average count rate measured: 386 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.82 +/- 0.01 e-/ADU RON = 6.79 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.161 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay