*** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 23 Feb 2002 13:03:09 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 23 Feb 2002 13:03:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020223 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.07 stddev = 2.29 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_020223 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1779.9 1777.6 629.5 0.00 -0.03 2.83 2 8.63 3502.0 3503.0 1243.1 0.01 0.08 2.82 3 17.25 6936.3 6937.0 2443.1 -0.01 0.01 2.84 4 25.87 10371.7 10370.9 3668.4 -0.01 -0.01 2.83 5 38.80 15521.9 15523.7 5438.9 -0.02 -0.01 2.85 6 51.73 20674.8 20675.5 7262.9 -0.02 -0.02 2.85 7 64.66 25828.3 25827.9 9077.0 -0.01 -0.02 2.85 8 86.21 34417.5 34420.9 12002.9 -0.01 -0.01 2.87 9 107.76 43017.8 43015.6 15139.6 0.02 0.00 2.84 10 129.31 51616.3 51612.3 18098.4 0.04 0.02 2.85 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5075 to 147266 e- Average count rate measured: 399 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.01 e-/ADU RON = 6.54 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.153 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay