*** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 12 Feb 2002 13:07:15 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 12 Feb 2002 13:07:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 563.46 stddev = 2.39 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_020212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1777.8 1776.7 635.7 0.00 -0.01 2.80 2 8.63 3496.4 3495.9 1228.7 -0.01 0.00 2.85 3 17.25 6927.7 6928.1 2461.6 0.02 0.03 2.81 4 25.87 10354.9 10357.3 3661.8 -0.01 0.01 2.83 5 38.80 15501.3 15498.8 5406.5 0.00 -0.02 2.87 6 51.73 20640.5 20643.0 7249.6 -0.03 -0.02 2.85 7 64.66 25787.7 25788.2 8996.2 -0.01 -0.02 2.87 8 86.21 34361.4 34365.1 12041.2 -0.01 -0.01 2.85 9 107.76 42951.7 42950.6 15372.6 0.03 0.02 2.79 10 129.31 51528.9 51527.1 17895.1 0.03 0.02 2.88 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5067 to 146908 e- Average count rate measured: 398 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.85 +/- 0.02 e-/ADU RON = 6.81 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.156 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay