*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Dec 2007 17:50:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Dec 2007 17:50:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 071227 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 605.57 stddev = 4.09 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_071227 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1896.7 1899.8 709.7 0.02 0.06 2.67 2 8.41 3722.4 3721.3 1347.4 0.04 -0.04 2.76 3 16.81 7365.3 7365.4 2680.6 -0.07 -0.08 2.75 4 25.21 11009.8 11010.0 3946.2 -0.09 -0.09 2.79 5 37.82 16485.3 16484.2 5940.9 -0.08 -0.08 2.77 6 50.42 21966.6 21965.7 7937.1 -0.03 -0.02 2.77 7 63.03 27456.4 27454.5 9955.7 0.02 0.02 2.76 8 84.04 36604.3 36607.1 13348.1 0.07 0.09 2.74 9 105.05 45728.9 45732.7 16336.0 0.05 0.07 2.80 10 126.05 54866.3 54861.8 19685.5 0.07 0.07 2.79 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5289 to 152873 e- Average count rate measured: 434 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.79 +/- 0.01 e-/ADU RON = 11.39 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.158 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay