*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Nov 2006 12:23:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Nov 2006 12:23:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061117 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.13 stddev = 3.50 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_061117 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 2111.6 2112.0 783.6 0.05 0.07 2.69 2 8.41 4139.8 4138.2 1505.6 -0.03 -0.06 2.75 3 16.81 8194.3 8195.9 2986.2 -0.09 -0.07 2.74 4 25.21 12248.6 12249.2 4487.1 -0.11 -0.11 2.73 5 37.82 18346.2 18344.3 6628.9 -0.07 -0.07 2.77 6 50.42 24447.3 24448.7 8909.1 -0.01 -0.00 2.74 7 63.03 30565.7 30562.0 11221.8 0.06 0.06 2.72 8 84.04 40732.3 40726.6 14684.7 0.07 0.06 2.77 9 105.05 50911.0 50899.6 18409.8 0.10 0.08 2.77 10 126.05 61032.2 61038.9 22047.7 0.03 0.05 2.77 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5850 to 169089 e- Average count rate measured: 483 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.77 +/- 0.01 e-/ADU RON = 9.70 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.156 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay