*** MIDAS logfile (05SEPpl1.2) *** Tu, 04 Jul 2006 17:40:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 04 Jul 2006 17:41:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060704 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.39 stddev = 4.25 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_060704 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 2181.4 2183.4 802.9 0.06 0.07 2.72 2 8.41 4286.7 4289.2 1568.3 -0.06 -0.04 2.73 3 16.81 8501.4 8501.8 3098.6 -0.08 -0.09 2.74 4 25.21 12713.1 12713.2 4608.8 -0.10 -0.11 2.76 5 37.82 19048.6 19044.7 6923.0 -0.06 -0.08 2.75 6 50.42 25390.8 25387.2 9352.2 0.01 0.00 2.71 7 63.03 31738.1 31737.7 11552.7 0.06 0.06 2.75 8 84.04 42300.5 42303.6 15247.3 0.07 0.08 2.77 9 105.05 52852.9 52857.2 18984.7 0.06 0.07 2.78 10 126.05 63392.4 63388.5 22994.0 0.04 0.04 2.76 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6047 to 175648 e- Average count rate measured: 502 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.77 +/- 0.01 e-/ADU RON = 11.79 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.133 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay