*** MIDAS logfile (05SEPpl1.2) *** We, 25 Jan 2006 12:34:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 25 Jan 2006 12:35:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060125 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 607.87 stddev = 6.35 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_060125 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1257.0 1255.1 496.9 0.05 0.02 2.53 2 8.41 2466.8 2467.0 925.7 -0.02 0.03 2.66 3 16.81 4886.8 4886.6 1755.3 -0.06 -0.04 2.78 4 25.21 7305.9 7305.6 2692.1 -0.08 -0.08 2.71 5 37.82 10938.3 10937.9 4008.4 -0.09 -0.09 2.73 6 50.42 14573.6 14575.5 5241.7 -0.05 -0.05 2.78 7 63.03 18213.7 18211.5 6643.5 -0.02 -0.04 2.74 8 84.04 24285.0 24288.8 8738.6 0.04 0.04 2.78 9 105.05 30365.7 30368.2 10922.2 0.10 0.10 2.78 10 126.05 36436.2 36433.6 13002.2 0.13 0.11 2.80 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3520 to 102111 e- Average count rate measured: 288 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.80 +/- 0.01 e-/ADU RON = 17.80 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.144 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay