*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 20:29:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 20:30:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060114 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 610.30 stddev = 5.44 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1256.3 1257.9 482.1 0.00 0.05 2.61 2 8.41 2473.0 2471.5 924.0 0.07 -0.02 2.68 3 16.81 4898.4 4897.9 1811.2 -0.05 -0.08 2.70 4 25.21 7326.1 7326.2 2664.5 -0.07 -0.07 2.75 5 37.82 10968.0 10971.1 3988.9 -0.10 -0.07 2.75 6 50.42 14616.7 14615.6 5285.3 -0.05 -0.05 2.77 7 63.03 18264.5 18264.9 6534.1 -0.04 -0.03 2.80 8 84.04 24356.5 24357.4 8858.6 0.03 0.04 2.75 9 105.05 30449.6 30451.4 10976.9 0.08 0.09 2.77 10 126.05 36542.3 36540.8 13361.3 0.11 0.12 2.73 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3466 to 100736 e- Average count rate measured: 289 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.76 +/- 0.01 e-/ADU RON = 14.98 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.135 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay