*** MIDAS logfile (05SEPpl1.2) *** Sa, 07 Jan 2006 16:21:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 07 Jan 2006 16:22:32 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060107 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.12 stddev = 3.46 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_060107 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1248.6 1249.9 466.6 0.03 0.08 2.68 2 8.41 2452.3 2450.6 895.3 0.01 -0.08 2.74 3 16.81 4857.8 4857.0 1758.9 -0.04 -0.06 2.76 4 25.21 7261.9 7261.4 2625.9 -0.08 -0.08 2.77 5 37.82 10874.7 10875.1 3916.2 -0.07 -0.05 2.78 6 50.42 14485.6 14483.2 5278.4 -0.05 -0.06 2.74 7 63.03 18102.5 18103.6 6550.5 -0.03 -0.01 2.76 8 84.04 24137.6 24135.2 8769.6 0.03 0.03 2.75 9 105.05 30180.0 30179.3 11061.2 0.09 0.10 2.73 10 126.05 36213.1 36211.7 13030.6 0.12 0.13 2.78 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3447 to 99932 e- Average count rate measured: 287 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.76 +/- 0.01 e-/ADU RON = 9.55 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.146 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay