*** MIDAS logfile (03SEPpl1.2) *** Su, 05 Jun 2005 13:59:17 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 05 Jun 2005 14:00:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 050605 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.40 stddev = 3.26 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_050605 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1325.3 1326.8 493.1 0.00 0.05 2.69 2 8.41 2605.6 2604.1 955.3 0.04 -0.04 2.73 3 16.81 5160.3 5162.1 1856.3 -0.05 -0.02 2.78 4 25.21 7716.3 7714.8 2808.2 -0.06 -0.08 2.75 5 37.82 11554.0 11550.9 4189.3 -0.07 -0.09 2.76 6 50.42 15390.8 15388.9 5575.2 -0.06 -0.06 2.76 7 63.03 19233.0 19233.3 6980.9 -0.04 -0.02 2.76 8 84.04 25647.7 25646.7 9323.6 0.03 0.04 2.75 9 105.05 32067.3 32066.6 11606.4 0.09 0.10 2.76 10 126.05 38473.1 38467.6 13778.5 0.10 0.10 2.79 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3686 to 106937 e- Average count rate measured: 305 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.78 +/- 0.01 e-/ADU RON = 9.06 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.143 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay