*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 26 Oct 2003 15:51:27 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 26 Oct 2003 15:52:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 031026 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 609.32 stddev = 3.15 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_031026 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1537.1 1536.6 572.0 0.06 0.01 2.69 2 8.41 3015.3 3018.2 1110.0 -0.06 0.03 2.72 3 16.81 5973.6 5976.0 2192.4 -0.09 -0.06 2.72 4 25.21 8933.0 8932.1 3203.3 -0.08 -0.10 2.79 5 37.82 13376.6 13377.8 4810.1 -0.07 -0.08 2.78 6 50.42 17821.9 17822.0 6655.3 -0.04 -0.05 2.68 7 63.03 22277.7 22276.9 8243.9 0.01 -0.00 2.70 8 84.04 29704.4 29703.8 10613.7 0.07 0.06 2.80 9 105.05 37123.3 37129.8 13500.8 0.09 0.09 2.75 10 126.05 44542.2 44542.8 15926.6 0.11 0.10 2.80 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4278 to 124000 e- Average count rate measured: 353 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.78 +/- 0.02 e-/ADU RON = 8.77 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.147 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay