*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Jul 2003 17:23:04 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Jul 2003 17:23:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 030706 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.50 stddev = 2.94 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_030706 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1583.8 1584.4 587.4 0.04 0.06 2.70 2 8.41 3110.6 3109.6 1123.1 0.02 -0.02 2.77 3 16.81 6158.2 6160.1 2272.1 -0.09 -0.06 2.71 4 25.21 9207.9 9207.6 3369.7 -0.10 -0.11 2.73 5 37.82 13790.2 13789.8 5028.9 -0.08 -0.08 2.74 6 50.42 18371.4 18374.0 6667.0 -0.06 -0.04 2.76 7 63.03 22968.2 22965.7 8432.7 0.01 0.00 2.72 8 84.04 30626.5 30623.6 11157.2 0.07 0.07 2.74 9 105.05 38274.0 38277.5 13753.5 0.09 0.10 2.78 10 126.05 45920.2 45916.6 16664.0 0.10 0.09 2.76 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4381 to 126991 e- Average count rate measured: 364 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.77 +/- 0.01 e-/ADU RON = 8.14 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.145 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay